Test Principle
The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Interleukin 6 (IL6). Standards or samples are then added to the appropriate microtiter plate wells with a biotin-conjugated antibody specific to Interleukin 6 (IL6). Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Interleukin 6 (IL6), biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Interleukin 6 (IL6) in the samples is then determined by comparing the O.D. of the samples to the standard curve.
Information
Research AreaCytokine Tumor immunity Infection immunity Cardiovascular biology
Species ReactivityRhesus monkey (Simian)
Sample Typeserum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Specifications
Tested ApplicationELISA
SensitivityThe minimum detectable dose of this kit is typically less than 5.8pg/mL
Detection Range15.6-1000pg/mL
Assay Time3h
Detection MethodEnzyme-linked immunosorbent assay for Antigen Detection.